2. Tools, Maintenance Schedules, Procedures and Testing
96-108639 Rev 3.0
Setting up the pressure fittings for use with the recommended pressure
reference meter(s)
NOTE: The pressure fitting (left hand fitting when facing the front of the cassette) is fitted with an
adaptor to attach ¼" BSP devices such as hypodermic pressure fittings. (See Figure 9 below).
This adaptor is not used with the standard SciCan pressure tube and should be removed from the
cassette extension fitting before attaching the pressure tube.
Figure 9. Pressure fitting cut-away
To set up the cassette for use with the recommended Druck DPI705R and Heise PTE1 pressure
reference meters, proceed as follows:
Remove the 1/8 NPT to ¼ BSP adaptor. (The accessories for the cassette include 2 male
Swagelok 'Quick connect' fittings (see Figure 10).
STAT IM 2000/2000S
STATIM 2000/2000S/2000 G4 Service Guide
Service Guide
Page 23