3. Troubleshooting Cycle Faults
To activate NVRAM backup, proceed as follows:
1. The UNWRAPPED or WRAPPED keys will both switch the screen display to "YES".
2. The RUBBER & PLASTICS key will action your selection and return the unit to the service menu
condition.
3. Pressing the STOP key will exit without changes and return the unit to the operational menu.
>Restore NVRAM
>Restore NVRAM
No
Screen display when in Restore NVRAM sub menu.
This function will allow the stored (backed up) unit settings to be restored if operational settings
are lost for any reason.
The default setting on all STATIM units is NO.
The alternative is YES.
To activate NVRAM restoration, proceed as follows:
1. The UNWRAPPED or WRAPPED keys will both switch the screen display to "YES".
2. The RUBBER & PLASTICS key will action your selection and return the unit to the service menu
condition.
3. Pressing the STOP key will exit without changes and return the unit to the operational menu.
>Save and Exit
Allows any changes made in the sub menus (where the STOP button does not action save) to be
saved and the unit will then exit to the operational menu.
To activate save and exit, press the RUBBER & PLASTICS button.
>Exit
Allows you to exit the service menu WITHOUT saving changes other than those changes saved by
activating the STOP button.
To activate exit, press the RUBBER & PLASTICS button.
>Water Pump Type
>Water Pump Type
Ulka
Screen display when in Press. Sensor Type sub menu.
This sub menu is for information only.
Page 62
STAT IM 2000/2000S
STATIM 2000/2000S/2000 G4 Service Guide
96-108639 Rev 3.0