SciCan STATIM 2000 Service Manual page 90

Hide thumbs Also See for STATIM 2000:
Table of Contents

Advertisement

4. STAT IM 2000 Calibration
11. Press the STOP button on the control box to end the steam generator calibration cycle
12. Press STOP a second time to reset the unit.
3. Chamber Thermocouple Calibration
To calibrate the chamber thermocouple:
1. Turn the power switch to OFF. Ensure the calibration jumper is in place and turn the power
2. To start a chamber calibration cycle, press and release the UNWRAPPED cycle button
3. Wait for the chamber to reach the sterilization temperature of 134 °C.
4. The temperature displayed on the LCD needs to match the temperature displayed on the
5. To change the temperature display on the LCD, adjust the TC-ADJ potentiometer (9)
6. Observe the temperatures displayed on the LCD and the digital thermometer for 30
7. Press the STOP button on the control box when finished. Wait for the beeps to signal that
4. Steam Generator Thermocouple Re-calibration
Repeat the steam generator thermocouple calibration procedure in section 2, Steam Generator
Thermocouple Calibration , above.
5. Switch off the unit and remove all calibration equipment.
Page 88
and to depressurize the cassette.
switch ON. The LCD will read (for example):
on the control box and then press the START button. The STAT IM will run a regular
UNWRAPPED cycle, but continue to show calibration information on the LCD.
digital thermometer.
on the Controller Board. Continue adjusting TC-ADJ until the temperatures are within
±0.25°C.
seconds. The temperatures registered should remain within ± 0.25°C of one another.
Adjust TC-ADJ if not.
the cassette may be removed.
STAT IM 2000/2000S
E8>>CALIBRATING*<<F5
+60.00°C 90.00°C
STATIM 2000/2000S/2000 G4 Service Guide
96-108639 Rev 3.0