SciCan STATIM 2000 Service Manual page 174

Hide thumbs Also See for STATIM 2000:
Table of Contents

Advertisement

8. Water Pumps, Reservoir, and Compressor
5. Inspect the filter and remove any debris. Rinse the filter in clean water.
6. Insert the filter and reassemble the inlet fitting to the pump body inlet fitting. Ensure that the
rubber seal is in place. Finger tighten the fitting and using a 9/16-inch wrench, tighten one half
turn ONLY so as not to damage the internal thread of the plastic pump fitting.
7. Using a 7/16-inch wrench, disconnect the right angle fitting (6) and collar sub assembly while
holding the pump body outlet fitting (11) with a 9/16-inch wrench and clean the threads of any
debris. Note the orientation of the fitting relative to the steam generator (12).
Tip:
correct orientation when re-assembling.
8. Clamp the right angled filter fitting using an adjustable wrench or a vice if available, and
unscrew the collar with a 9/16-inch wrench.
9. There is an insert (13) containing a fine mesh filter in the end of the right angle fitting. This filter
cannot be removed from the fitting. If there are any deposits on the mesh, clean off with a
stiff, non-metallic brush. If available, an air line can be used to purge the debris from the filter
if appropriate. If there is any scaling on the fitting, it can be removed by soaking the fitting
and filter in a mild solution suitable for removing scale or mineral deposits (i.e. vinegar). Rinse
thoroughly with clean water.
10. If the outlet fitting has become damaged beyond use, then the in-line filter should be removed,
and the outlet fitting replaced with replacement fitting 01-111115S, which has no filter. The
unit should be upgraded with an in-line filter and reservoir filter, if appropriate.
11. If replacing the outlet fitting or re-fitting when the components are clean, apply a small
amount of Teflon™ tape or pipe thread compound on to the external thread of the fitting
and reassemble the right angle fitting to the collar ensuring the correct orientation of the two
components.
Page 172
Draw a black line through the three parts on the output side of the pump to ensure
Filter fitting and collar assembly (dis-assembly)
STAT IM 2000/2000S
STATIM 2000/2000S/2000 G4 Service Guide
96-108639 Rev 3.0
Figure 7
Figure 8
Service Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Statim 2000sStatim 2000 g4

Table of Contents