9. Solenoid Valve and Pressure Transducer
Page 192
WARNINGS AND
PRECAUTIONS
If you have questions about the unit you are repairing, please do not hesitate to
contact your local SciCan representative for information. Also, the STAT IM is heavy.
Exercise caution and seek assistance when lifting or carrying units.
EXERCISE CAUTION
Hazardous voltages are accessible when the cover is removed.
Disconnect the power cord before servicing the power mains portion of the
controller board and associated devices.
PERFORM TESTS
If the cover is removed, a dielectric strength test (Hi-Pot) AND a protective
bonding impedance test (ground continuity) must be performed on the STAT IM
when the work is completed and after the cover has been returned to the unit.
PROTECT THE UNIT
Use only steam-process distilled water in the STAT IM .
The STAT IM contains electronic circuitry that is static sensitive. Always wear a
static strap when working with or near printed wiring boards. In addition, use
static footstraps, grounding mats and grounded work surfaces when servicing
microprocessor devices. Transport boards and devices in static protected bags.
Ensure there is sufficient steam-process distilled water in the STAT IM before
activating the pump.
In order to ensure adherence to the applicable safety agency approvals,
state, provincial, regional and national laws, replace components with SciCan
approved parts only.
STAT IM 2000/2000S
STATIM 2000/2000S/2000 G4 Service Guide
96-108639 Rev 3.0
Service Guide