2. Tools, Maintenance Schedules, Procedures and Testing
Assembling the Swagelok B-QC4-S-2PM 'Quick Connect' fitting to the internal
pressure module supplied with the Heise hand held calibrator.
NOTE: You do NOT need the 1/8 NPT female to ¼ NPT male adaptor for this device. (see Figure
17 for assembly).
Insert and tighten the Swagelok fitting into the 1/8 NPT female thread of the pressure module. DO
NOT OVERTIGHTEN FITTING. NOTE: The use of P.T.F.E. tape as shown is essential to ensure
the connection is free from steam leaks. If the joint leaks, steam can enter the module and damage
the unit internally.
The pressure tube can now be attached to the pressure module.
STAT IM 2000/2000S
STATIM 2000/2000S/2000 G4 Service Guide
Service Guide
96-108639 Rev 3.0
Figure 17. Heise meter with
Swagelok fitting attached to
pressure module.
Page 29