12. Printer and Data Logger
7. Insert USB flash drive or SD memory card.
8. After a few seconds the LCD display should now show the following:
HH:MM DD/MM/YYYY
USB/FLASH DETECTED/SAFELY
REMOVE MSD/SELECT A CYCLE
Messages "USB/FLASH DETECTED", "SAFELY REMOVE MSD" and "SELECT A CYCLE" will
rotate in sequence and not all appear at the same time.
9. The unit is now ready for use.
Installation on 25 pin units (2000 ONLY).
(Note: these units will NOT have a user and service menu)
For this operation the 25 to 25 pin special cable supplied with the data logger is required.
Note that this cable is directional and can only be fitted one way. The cable sockets are
marked "STATIM" and "data logger" to assist in identifying the direction of connection.
Failure to connect the cable in the correct direction will cause the LCD screen to go blank,
and may cause damage to the unit.
This type of STATIM 2000 unit (no menu facility) will automatically recognize the data logger when
connected. To install the data logger, proceed as follows:
1. Connect the Data Logger to the power cable attached to the transformer supplied. Important
note: ensure that the plug is fully inserted in the socket on the data logger.
2. Attach the adaptor suitable for your regional power supply on to the main body of the
transformer. Plug in to the mains supply BUT DO NOT SWITCH ON AT THIS TIME.
3. Ensure that the STATIM unit is powered OFF.
4. Using the 25 pin to 25 pin cable supplied, connect the Data Logger to the STATIM unit and
secure the cable. Note the orientation of the cable.
STAT IM 2000/2000S
STATIM 2000/2000S/2000 G4 Service Guide
Service Guide
96-108639 Rev 3.0
Page 253