4. STAT IM 2000 Calibration
2. Validation thermocouple calibration
To calibrate the validation thermocouple:
1. Turn the unit off OFF.
2. Turn ON the unit while keeping Unwrapped and Wrapped button pressed to enter STAT IM
3. This Service Mode is password protected, enter password to continue (default password
4. Toggle using keypad through the menu selection to reach Calibration option and press
5. The display should show the calibration screen as before
6. To start a validation thermocouple (self-calibrating) cycle, press and hold the
7. If the asterisk "*" does not appear, press the STOP button on the control box twice to
8. Hold down the UNWRAPPED cycle button on the control box and press the START button
9. This calibration will take approximately 6 minutes.
10. Allow the self-calibration to complete.
11. The temperature within the chamber will rise to the sterilization temperature. Wait until
12. The offset value in the upper left-hand corner of the display (CVO) may have changed to a
13. Press the STOP button on the control box when finished.
3. Switch off the unit and remove all calibration equipment.
Page 96
Service Mode.
is: Unwrapped, Wrapped, R&P and Stop keys pressed in this order).
R&P key.
UNWRAPPED cycle button on the keypad/control box and press the START button. An
asterisk "*" will appear next to the validation thermocouple offset value (CVO) to indicate
that a Validation Assy. calibration cycle is running.
reset the unit. Turn the power switch OFF then ON again.
to start the self-calibration again.
sterilization phase of the calibration cycle ends automatically.
new offset value.
STAT IM 2000/2000S
STATIM 2000/2000S/2000 G4 Service Guide
96-108639 Rev 3.0
Service Guide