5. STAT IM 2000S Calibration
7. Turn ON the unit while keeping Unwrapped and Wrapped button pressed to enter STAT IM
Service Mode.
8. This Service Mode is password protected, enter password to continue (default password is:
Unwrapped, Wrapped, R&P and Stop keys pressed in that order).
Keypad functions (Unit LCD or Control Box) at this time will be:
Unwrapped Key: Select next item in the menu.
Wrapped Key: Select previous item in the menu.
Rubber and Plastics Key: Enter current selection
Toggle using keypad through the menu selection to reach Calibration option and press
R&P key.
The display should be similar to the diagram below.
Where
PPP.P = actual chamber pressure in kPa.
PO = chamber pressure hexadecimal offset value
TTT.T = theoretical pressure calculated from the actual chamber temperature.
CO = chamber temperature hexadecimal offset value
CCC.C = actual chamber temperature in °C
IMPORTANT NOTES:
Remember – the steam generator thermocouple does NOT require calibration.
The VREF does not need adjustment.
STAT IM 2000/2000S
STATIM 2000/2000S/2000 G4 Service Guide
PPP.P PO TTT.T
CO
CCC.C
Service Guide
96-108639 Rev 3.0
Page 115