Hygiene instrument, reprocessing container
Hide thumbs Also See for SALUS:

Advertisement

Quick Links

Operator's Manual
SALUS
®
- Hygiene Instrument
Reprocessing Container

Advertisement

Table of Contents
loading

Summary of Contents for SciCan SALUS

  • Page 1 Operator’s Manual SALUS ® - Hygiene Instrument Reprocessing Container...
  • Page 2: Table Of Contents

    HOW TO CALCULATE THE END OF LIFE EXPIRY DATE USING THE CONTAINER Opening the container: Preparing and loading instruments Load instruments REPROCESSING SALUS – HYGIENE SLEEVE AND RACK WITH INSTRUMENTS After use - used instruments Cleaning the Rack with Instruments 8.2.1 In a medical washer or thermal disinfector 8.2.2...
  • Page 3: Indications For Use

    DISCLAIMERS Do not permit any person other than certified personnel to supply parts for, service or maintain your SALUS - Hygiene. SciCan will not be liable for incidental, special or consequential damages caused by any maintenance by a third party including lost profits, any commercial loss, economic loss, or loss arising from personal injury.
  • Page 4: Overview

    Accessories Spare parts contents Air/water syringe tip clip CASSETTE-CLIP Gasket seal, front 01-114777S When you receive your SALUS - Hygiene packing carton, the items Labels (English/Spanish, 100/pack) 01-114360 Gasket seal, back 01-114778S listed below will be included. If any of...
  • Page 5: Specifications

    Example: A dental office runs 4 sterilization cycles a day and is open 150 days of the year, in 4 years SALUS - Hygiene will need to be retired. Upon the filter label and the rack label, you would write today’s date in the date of service box, add four years and write that date in the “end of life”...
  • Page 6: Using The Container

    Before loading any instruments into the SALUS - Hygiene, consult the instrument manufacturer’s reprocessing instructions. IMPORTANT: The SALUS - Hygiene is not designed to allow cleaning of instruments with lumens or channels. IMPORTANT: The SALUS - Hygiene has not been validated to process implantable medical devices.
  • Page 7: Cleaning The Rack With Instruments

    Cleaning the Rack with Instruments Unlock the container latch and separate the rack from the sleeve. Remove any large debris from the rack. 8.2.1 In a medical washer or thermal disinfector Place the rack with instruments in an automated washer or thermal disinfector and reprocess according to the medical washer or thermal disinfector manufacturer’s instructions.
  • Page 8: Preparing For Sterilization

    CAUTION: Not all chemical indicators are compatible with the SALUS – Hygiene due to their shapes, sizes and pass/fail indication. Make sure the section of the chemical indicator allows for clear and unambiguous interpretation of its result through the transparent container sleeve.
  • Page 9: Transporting And Storing The Container

    Burns from hot surfaces. When removing the SALUS - Hygiene after a sterilization cycle, exercise caution to avoid skin burns as the container surfaces may still be hot. Allow SALUS - Hygiene to cool sufficiently before handling or wear heat-resistant gloves.
  • Page 10: Storing

    Storing SALUS - Hygienes containers can be stacked in storage and the sterility of the container contents is ensured for 30 days. SALUS - Hygiene containers should be stored in a manner that reduces the potential for contamination. Recommended storage conditions for SALUS- Hygiene are between 64 and 75 ºF (18 and 24 ºC) with relative humidity between 30 to 70%.
  • Page 11: Reusable Filter Inspection And Care

    11.2 Seal inspection and care The SALUS - Hygiene has a gasket seal on both sides (front and rear) of the rack. Inspect the gasket seals to ensure they are free of tears, deterioration, debris and indentation. A damaged seal must be replaced.
  • Page 12: Manufacturer's Information And Customer Service Information

    For all service and repair inquiries: In Canada: 1-800-870-7777 United States: 1-800-572-1211 Germany: +49 (0) 7561-983-620 International: (416) 446-4500 techservice.ca@scican.com Manufactured by: SciCan Ltd. 1440 Don Mills Road, Toronto, ON M3B 3P9 Canada Telephone: (416) 445-1600 Fax: (416) 445-2727 Toll free: 1-800-667-7733 SciCan Inc.
  • Page 13: Warranty

    SciCan’s option and without charge of any defected part(s), provided SciCan is notified in writing within thirty (30) days of the date of such a failure and further provided that the defective part(s) are returned to SciCan prepaid.
  • Page 14: Salus - Hygiene Placement In Racks