6. Cover Assembly
3. Carefully remove any residual adhesive or membrane from the recessed areas of the fascia
using a sharp blade. DO NOT USE AROMATIC SOLVENTS TO REMOVE ADHESIVE FROM
THE FASCIA OR THE MEMBRANE KEYPAD.
To replace the membrane (1), proceed as follows (see Figure 3):
1. Before installing the membrane onto the fascia, connect the new keypad cable (3) connector
to controller board P4 (4), holding the keypad connector by its tab, insert it into the housing on
the controller board and pull up on the tab to lock the connector into place.
2. Plug in the power cord and power the unit ON. To test each button on the keypad, ensure that
a cassette is not inserted into the armature. Select a cycle and press the START button. The
LCD will display "INSERT CASSETTE". Repeat this step for each cycle to ensure the keypad is
functioning properly. To test the STOP button, select a cycle. The LCD message will indicate
which cycle is selected. Press the STOP button. The LCD message will change to "SELECT A
CYCLE".
3. Turn off and unplug the unit, and disconnect the keypad cable, by gently releasing the lock
on the connector housing of the controller board and pulling down on the connector tab to
release it.
4. The replacement membrane keypad has carrier paper on the back to protect the adhesive.
With the carrier paper in place, feed the keypad cable (4) connector through the slot (5) in the
fascia (2).
5. Remove the carrier paper and align the bottom edge of the membrane keypad with the bottom
edge of the recessed area on the fascia. Carefully lower the membrane into place while
continuing to draw the keypad cable connector through the slot. Using finger pressure only,
press the membrane into place, eliminating air pockets under the membrane.
6. Secure the keypad cable and LCD cable together using double-sided adhesive tape.
7. Connect the keypad cable (3) connector to controller board P4 (4) holding the keypad
connector by its tab, insert it into the housing on the controller board and pull up on the tab to
lock the connector into place.
8. Connect the LCD cable (6) connector to controller board (7).
9. Test the membrane keypad again by repeating steps 1 and 2.
Page 128
STAT IM 2000/2000S
STATIM 2000/2000S/2000 G4 Service Guide
96-108639 Rev 3.0