5. STAT IM 2000S Calibration
7. Turn ON the unit and access the service menu in the following manner:
8. This Service Mode is PIN protected, enter password to continue (default password is: 7919).
9. From the service menu, select Calibration.
IMPORTANT NOTES:
Remember – the steam generator thermocouple does NOT require calibration.
The VREF does not need adjustment.
1. Chamber Thermocouple and Pressure Transducer Calibration
To calibrate the chamber thermocouple and pressure transducer:
1. To start the calibration cycle, select Calibration in the Calibration submenu and calibration will
start automatically
STAT IM 2000/2000S