Advertisement

4. STAT IM 2000 Calibration
3. Validation thermocouple calibration
To calibrate the validation thermocouple:
4. Switch off the unit and remove all calibration equipment.
Page 92
Turn the power switch to OFF. Ensure the calibration jumper is in place and turn the power
switch ON.
To start a validation thermocouple (self-calibrating) cycle, press and hold the
UNWRAPPED cycle button on the control box and press the START button. An asterisk
* will appear next to the validation thermocouple offset value (CVO) to indicate that a
validation assembly calibration cycle is running.
If the asterisk * does not appear, press the STOP button on the control box twice to reset
the unit. Turn the power switch OFF then ON again.
Hold down the UNWRAPPED cycle button on the control box and press the START button
to start the self-calibration again.
This calibration will take approximately 6 minutes.
Allow the self-calibration to complete.
The temperature within the chamber will rise to the sterilization temperature. Wait until
sterilization phase of the calibration cycle ends automatically.
The offset value in the upper left-hand corner of the display (CVO) may have changed to a
new offset value.
Press the STOP button on the control box when finished.
STAT IM 2000/2000S
STATIM 2000/2000S/2000 G4 Service Guide
96-108639 Rev 3.0
Service Guide