Faults, Cause and Remedy

Faults, Cause and Remedy
Fault
no output signal
transmitter output signal and central
device display do not match
output signal 1 mA
high linearity error
possible calibration range at central
device exhausted
20
Cause
gas transmitter is not powered up
gas transmitter defective
central device and gas transmitter are
not matched
ambient temperature too high resp. too
low
gas transmitter defective
wrong target gas category set
calibration range at central device too
small
Remedy
Check power supply and polarity.
Have Dräger Safety check the gas trans-
mitter.
Match central device and gas transmitter,
see "Calibration" on page 12.
Operate gas transmitter within the speci-
fied temperature range, see "Technical
Data" on page 21.
Have Dräger Safety check the gas trans-
mitter.
Change the target gas category, see
"How to Change the Target Gas Cate-
gory" on page 17.
Calibrate system at gas transmitter.