Command 48 Status Bits
1066 ph Device variables
Byte / Bit
1 / 0
Cpu Error / Error
The software checksum is not as expected. The CPU memory has been
corrupted.
1 / 1
Self-Test Fail / Error
An electronic component is out of specification.
1 / 2
Factory Data Error / Error
An error was detected in the factory segment of the non-volatile memory.
At least one factory configuration parameter has been corrupted.
1 / 3
hardware/Software Mismatch / Error
The software is not compatible with the hardware.
1 / 4
Internal Communications Error / Error
The analog input electronics is non-responsive.
3 / 0
Keypad Error / Warning
At least one key in the device keypad is stuck. This condition makes the
local operator interface unuseable. If no other alerts are present, the
device can still perform its other functions normally.
3 / 1
user Data Error / Warning
An error was detected in the non-volatile memory. One or more user con-
figuration parameter may be corrupted. Reset analyzer to factory defaults
and re-configure the device. If the problem persists, replace device.
3 / 2
Need Factory Calibration / Warning
The device's non-volatile memory has been corrupted. The device meas-
urements may be out of specification.
3 / 3
Software Mismatch / Warning
The input CPU software is not fully compatible with the main CPU soft-
ware.
3 / 7
Reset In progress / Other
The transmitter's configuration is being reset to factory defaults.
6 / 6
Maintenance Required / Other
6 / 1
Device variable alert / Other
8 / 0
Simulation active / Mode
A device variable is being simulated.
8 / 1
Non-volatile Memory Defect / Warning
8 / 2
volatile Memory Defect / Error
8 / 3
Watchdog Reset Executed / Other
8 / 4
power Supply Condition Out Of Range / Warning
8 / 5
Environmental Condition Out Of Range / Warning
8 / 6
Electronic Defect / Error
Reset device or turn power off then on. If problem persists, replace device.
HART Communications
Meaning / Class
Section 10: haRT Communication
Device Status Bits Set
4 – More Status available
7 – Field Device Malfunction
4 – More Status available
7 – Field Device Malfunction
4 – More Status available
7 – Field Device Malfunction
4 – More Status available
7 – Field Device Malfunction
4 – More Status available
7 – Field Device Malfunction
4 – More Status available
4 – More Status available
4 – More Status available
4 – More Status available
4 – More Status available
4 – More Status available
4 – More Status available
4 – More Status available
4 – More Status available
4 – More Status available
4 – More Status available
4 – More Status available
4 – More Status available