1066 Temperature parameters
rtd_offset
The temperature offset resulting from a temperature calibration.
rtd_slope
The slope (dimensionless) resulting from a two point temperature
calibration. (Valid for cell constants < 0.02 1/cm)
temp_comp_mode
Selects automatic or manual temperature compensation.
manual_temperature
If manual temperature compensation is chosen, provides a con-
stant temperature value used by the transmitter.
1066 ph/ORp
zero_offset
The zero offset of a pH sensor resulting from a calibration.
ph_slope
The slope of a pH sensor resulting from a buffer calibration.
zero_offset_limit
The maximum value of the zero offset allowed for a successful
calibration.
min_ph_slope
The minimum value of the slope allowed for a successful calibration.
max_ph_slope
The maximum value of the slope allowed for a successful calibration.
ph_stabilization_time
The period of time used to determine stability of a pH sensor during
automatic calibration.
ph_stabilization_value
The pH change used to determine stability of a pH sensor during
automatic calibration.
ph_buffer_standard
The types of pH buffers available for automatic calibration.
preamp_location
The location of the preamplifier in a pH measurement.
reference_impedance_level
Configures the transmitter to use reference electrodes with
impedances less than 500 kohm (low) or greater than 500 kohm
(high).
ph_solution_temperature_correction
Configures the transmitter to correct the pH measurement for changes
in the pH of the solution with temperature.
ph_solution_temperature_coefficient
The coefficient used to correct solution pH when using custom
solution pH temperature correction.
HART Communications
Section 10: haRT Communication
Type
Enumeration / Range
FLOAT
FLOAT
ENUM
Manual
Automatic
FLOAT
Type
Enumeration /
Range
FLOAT
FLOAT 40 -- 62 mV/pH R / W
FLOAT
FLOAT 40 -- 62 mV/pH R / W
FLOAT 40 -- 62 mV/pH R / W
FLOAT 0 -- 99 seconds
FLOAT 0.01 -- 1.0 pH
ENUM Standard / Nist
DIN 19267
Ingold
Merck
Fisher
ENUM analyzer
sensor / J-box
ENUM Low
High
ENUM Off
ultra pure water
high pH
ammonia
custom
FLOAT -9.999 -- 9.999
pH / °C
April 2017
Default
Read/Write
R / W
0
R / W
1
R / W
Automatic
R / W
25.0 °C
Default
Read/Write
R / W
0.0 mV
59.16 mV/pH
R / W
60
40 mV/pH
62 mV/pH
R / W
10 seconds
R / W
0.02 pH
R / W
Standard / Nist
R / W
R / W
Low
Off
R / W
0.000 pH / °C