Measure; Range; Cell Constant; Temp Comp - Emerson Rosemount 1066 Instruction Manual

Smart-enabled, 2-wire transmitter
Hide thumbs Also See for Rosemount 1066:
Table of Contents

Advertisement

Instruction Manual
LIQ-MAN-1066
8.5.3

Measure

The display screen for selecting the measurement is shown. The default
value is displayed in bold type. Refer to the toroidal conductivity
Programming flow diagram to complete this function.

Range

8.5.4
The display screen for Selecting Auto-ranging or a specific range is
shown. The default value is displayed in bold type. Note: Ranges are
shown as conductance, not conductivity. Refer to the toroidal
conductivity Programming flow diagram to complete this function.
NOTE: when manually changing ranges, a zero calibration and
recalibration in a solution of known conductivity must be
performed with the toroidal sensor wired to the instrument. Refer
to
Section 9.5.2
Calibrating the Sensor in a Conductivity Standard.

Cell Constant

8.5.5
The display screen for entering a cell Constant for the sensor is shown.
The default value is displayed in bold type. Refer to the toroidal
conductivity Programming flow diagram to complete this function.
NOTE: When manually changing ranges, the Cell Constant may
change through the calibration process.
8.5.6

Temp Comp

The display screen for Selecting Temperature Compensation as Slope,
Neutral Salt, or Raw is shown. The default value is displayed in bold type.
Refer to the toroidal conductivity Programming flow diagram to
complete this function.
Programming Measurements
zeroing the Instrument and
Section 8: programming Measurements
1.234 µS/cm
Resistivity
Conductivity
TDS
Salinity
NaOH (0-12%)
HCl (0-15%)
Low H2SO4
High H2SO4
NaCl (0-20%)
Custom Curve
1.234 µS/cm
Auto
50-600 µS
0.5-100 mS
90-1500 mS
Section 9.5.3
1.234 µS/cm
2.7000 /cm
1.234 µS/cm
Slope
Raw
April 2017
25.0 ºC
Measurement
25.0 ºC
SN Range
25.0 ºC
Cell Constant
25.0 ºC
Temp Comp
51