Instruction Manual
LIQ-MAN-1066
Ozone Measurement programming
8.8
This section describes how to configure the 1066 transmitter for ozone measurement using
amperometric ozone sensors. The following programming and configuration functions are
covered:
1. Units: ppm Select ppm, mg/L, ppb, µg/L
2. Resolution: 0.001 Select display resolution 0.01 or 0.001
3. Filter: 5sec Override the default input filter, enter 0-999 seconds
To configure Ozone:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
The adjacent screen format will appear (factory defaults are shown). To program any displayed
function, scroll to the desired item and press ENTER.
The following sub-sections show the initial display screen that appears for each configuration
function. Use the flow diagram for ozone programming at the end of Sec. 6 and the 1066 live
screen prompts for each function to complete configuration and programming.
Note: Ozone measurement boards are detected automatically by the analyzer. No measurement
selection is necessary.
8.8.1
Units
The display screen for selecting measurement units is shown. The default
value is displayed in bold type. Refer to the Ozone Programming flow dia-
gram to complete this function.
8.8.2
Filter
The display screen for entering the input filter value in seconds is shown.
The default value is displayed in bold type. Refer to the Ozone
Programming flow diagram to complete this function.
Resolution
8.8.3
The display screen for selecting display resolution as 0.001 or 0.01 is
shown. The default value is displayed in bold type. Refer to the Ozone
Programming flow diagram to complete this function.
Programming Measurements
Section 8: programming Measurements
1.234 µS/cm
Units:
Filter:
Resolution: 0.001
1.234 µS/cm
mg/L
ppm
ppb
µg/L
1.234 µS/cm
05 sec
1.234 µS/cm
0.01
0.001
April 2017
25.0ºC
SN Configure
ppm
5 sec
25.0 ºC
SN Units
25.0 ºC
SN Input filter
25.0 ºC
SN Resolution
59