Quick Start Guide 00825-0100-3166, Rev AA April 2019 ™ Rosemount 1066 Single Channel Transmitter...
Page 2
Emerson products. Failure to follow the proper instructions may cause any one of the following situations to occur: loss of life, personal injury, property damage, damage to this instrument, and warranty invalidation.
April 2019 Quick Start Guide WARNING Electrostatic ignition hazard Special conditions for safe use (when installed in hazardous areas) The plastic enclosure, except the front panel, must only be cleaned with a damp cloth. The surface resistivity of the non-metallic enclosure materials is greater than one gigaohm. Take care to avoid electrostatic charge build-up.
Page 4
Quick Start Guide April 2019 Emerson.com/Rosemount...
2. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Emerson immediately. Mount Figure 1-1: Panel Mount Front View Quick Start Guide...
Page 6
Figure 1-2: Panel Mount Side View A. 4X mounting brackets and screws provided with instrument B. Panel supplied by others. Maximum thickness: 0.375 in. (9.52 mm) C. Panel mount gasket Figure 1-3: Panel Mount Bottom View A. Conduit openings Emerson.com/Rosemount...
Page 7
April 2019 Quick Start Guide Figure 1-4: Panel Cut-Out A. Maximum Figure 1-5: Wall Mount Front View A. 2-in. (50.8 mm) pipe supplied by customer B. 4 cover screws Quick Start Guide...
Page 8
Quick Start Guide April 2019 Figure 1-6: Wall Mount Side View Figure 1-7: Pipe Mount Bottom View A. Front panel B. 2-in. (50.8 mm) pipe mount bracket C. 2 sets of U-bolts for 2-in. (50.8 mm) pipe in kit PN 23820-00 Emerson.com/Rosemount...
Page 9
April 2019 Quick Start Guide Figure 1-8: Pipe Mount Side View General installation information 1. Install the transmitter with a sunshield or out of direct sunlight and areas with extreme temperatures. 2. Install the system in an area where vibrations and electromagnetic and radio frequency interference are minimized or absent.
Page 10
April 2019 Prepare conduit openings There are six conduit openings in all configurations of the transmitter. Note Emerson fits four of openings with plugs upon shipment. Figure 1-9: Conduit openings A. Front panel/keypad B. Power leads C. Alarm relay leads D.
Page 11
April 2019 Quick Start Guide To maintain ingress protection for outdoor use, seal unused conduit holes with suitable conduit plugs. Note Use watertight fittings and hubs that comply with your requirements. Connect the conduit hub to the conduit before attaching the fitting to the transmitter.
The power/signal cable can also be enclosed in an earth-grounded metal conduit. Note Do not run power supply/signal wiring in the same conduit or cable tray with loop power lines. Keep power supply/signal wiring at least 6 ft. (2 m) away from heavy electrical equipment. Emerson.com/Rosemount...
Page 13
April 2019 Quick Start Guide ™ Figure 2-1: Rosemount 1066 system block diagram A. 4-20 mA and HART signal B. Control system C. Computer D. Bridge E. Field Communicator ® 2.3.1 HART power supply and load requirements Table 2-1: HART Power Supply Standard Units Minimum supply voltage at transmitter terminals...
Page 14
2.3.2 HART current output wiring Emerson ships all instruments with two 4-20 mA current outputs. Current output 1 is the HART communications channel. Current output 2 is available to report process temperature measured by the temperature sensing element or RTD within the sensor.
Page 15
A. Install plugs in all other openings as needed. B. Ground lugs C. Hinge side of front panel D. Rosemount 1066 HART circuit board (pH/amperometric) ASSY 24539-00 E. TB5/THUM terminal is only used for wireless THUM adaptor instllations. F. TB7/output 2 requires external DC power.
Page 16
Figure 2-4: Configuring Rosemount 1066P Transmitter with F OUNDATION Fieldbus A. DeltaV configurator and host B. Fieldbus technician configurator C. Other host D. Power supply E. Filter F. Terminator G. Process line H. pH sensor I. Terminator Emerson.com/Rosemount...
Page 17
Use integral cable SMART sensors or compatible VP8 pH cables to ™ wire the Rosemount SMART pH sensors to the Rosemount 1066. 2. After wiring the sensor leads, use wiring diagrams found in the sensor manual to guide you as you carefully take up the excess sensor cable through the cable gland.
Page 18
Quick Start Guide April 2019 Figure 2-5: pH/ORP Sensor Wiring to the Rosemount 1066 Printed Circuit Board ® A. Rosemount 1066 circuit board ASSY 24539-00 (HART Note A. If ground lead is present, terminate it to green ground screw on inner enclosure.
Page 19
April 2019 Quick Start Guide Figure 2-6: Contacting and Toroidal Conductivity Sensor Wiring to the Rosemount 1066 Circuit Board A. Rosemount 1066 circuit board ASSY 24638-00 (HART) Quick Start Guide...
Page 20
Quick Start Guide April 2019 Figure 2-7: Chlorine, Oxygen, Ozone Sensor Wiring to Rosemount 1066 Printed Circuit Board A. Rosemount 1066 circuit board ASSY 24406-xx Note A. TB1, TB4, TB6, and TB7 not used for oxygen and ozone sensor wiring.
Page 21
Figure 2-8: Power/current loop wiring with wireless THUM adaptor A. Install plugs in all other openings as needed. B. Hinge side of front panel C. Rosemount 1066 HART circuit board (pH/chlorine/dissolved oxygen/ ozone) ASSY 24406-xx D. TB6/THUM terminal is used only for wireless THUM adapter installations.
Page 22
Figure 2-9: HART Loop Power Wiring A. Install plugs in all other openings as needed. B. Hinge side of front panel C. Rosemount 1066 HART circuit board (pH/chlorine/dissolved oxygen/ ozone) ASSY 24406-xx D. TB6/THUM terminal is used only for wireless THUM adapter installations.
April 2019 Quick Start Guide Navigating the display User interface The transmitter has a large display which shows the measurement readout and temperature in large digits and up to four additional process variables or diagnostic parameters concurrently. The displayed variables can be customized to meet user requirements.
Selection keys are used to: 1. Select items on the menu screens. 2. Scroll up and down the menu lists. 3. Enter or edit numeric values. 4. Move the cursor to the right or left. 5. Select measurement units during operation. Emerson.com/Rosemount...
Page 25
April 2019 Quick Start Guide Main display The transmitter displays the primary measurement value and temperature and up to four secondary measurement values, a fault and warning banner, and a digital communications icon. Process measurements One process variable and process temperature are displayed by default. For all configurations, the upper display area shows the live process variable, and the center display area shows the temperature (default screen settings).
This convenientlly allows display of the live values during important calibration and programming operations. Menu screens will time out after two minutes and return to the main live display. Emerson.com/Rosemount...
April 2019 Quick Start Guide Start up transmitter Procedure 1. Wire sensor(s) to the signal boards. Wire for wiring instructions. Refer to the sensor Quick Start Guide for additional details. Make current output, alarm relay, and power connections. 2. Once connections are secured and verified, apply DC power to the transmitter.
Page 28
Refer to the appropriate menu. 10. To return the transmitter to the factory default settings, choose Program under the main menu and then scroll to Reset. ™ Please call Rosemount Customer Support Center at 1-800-854-8527 if you need further support. Emerson.com/Rosemount...
April 2019 Quick Start Guide China RoHS table Quick Start Guide...
Page 32
Linkedin.com/company/Emerson- Automation-Solutions The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a twitter.com/rosemount_news mark of one of the Emerson family of companies. Facebook.com/Rosemount All other marks are the property of their youtube.com/RosemountMeasurement respective owners.