Recommended Sensors For Conductivity; Toroidal Conductivity (Codes - T); Performance Specifications - Emerson Rosemount 1066 Instruction Manual

Smart-enabled, 2-wire transmitter
Hide thumbs Also See for Rosemount 1066:
Table of Contents

Advertisement

Section 2: Description and Specifications
April 2017

Recommended Sensors for Conductivity

2.4.2
All Rosemount 400 series conductivity sensors (Pt 1000 RTD) and 410VP 4-electrode sensor.
Toroidal Conductivity (Codes
2.5
Measures conductivity in the range of 1 µS/cm to 2,000,000 µS/cm (2 S/cm). Measurement choices
are conductivity, resistivity, total dissolved solids, salinity, and % concentration. The % concentration
selection includes the choice of five common solutions (0-12% NaOH, 0-15% HCl, 0-20% NaCl, and 0-
25% or 96-99.7% H
temperature compensated. For other solutions, a simple-to-use menu allows the customer to enter
his own data. The transmitter accepts as many as five data points and fits either a linear (two points)
or a quadratic function (three to five points) to the data. Reference temperature and linear tempera-
ture slope may also be adjusted for optimum results. Two temperature compensation options are
available: manual slope (X% / °C) and neutral salt (dilute sodium chloride). Temperature compensa-
tion can be disabled, allowing the transmitter to display raw conductivity. For more information con-
cerning use of the toroidal conductivity sensors, refer to the product data sheets.

Performance Specifications

2.5.1
Temperature specifications:
Temperature range
Temperature Accuracy,
Pt-100, -25 to 50 °C
Temperature Accuracy,
Pt-100, 50 to 210 °C
TaBlE 2-2. performance Specifications: Recommended Range – Toroidal Conductivity
Model
1 S/cm
10 S/cm
226
225 & 228
242
222
(1in & 2in)
loop performance
(Following Calibration)
8
SO
). The conductivity concentration algorithms for these solutions are fully
2
4
-25 to 210 °C (-13 to 410 °F)
± 0.5 °C
± 1 °C
100 S/cm
1000 S/cm
50 S/cm to 500mS/cm
50 S/cm to 1500mS/cm
100 S/cm to 2000mS/cm
500 S/cm to 2000mS/cm
226: ±1% of reading ±5 S/cm in recommended range
225 & 228: ±1% of reading ±15 S/cm in recommended range
222, 242: ±4% of reading ±5mS/cm in recommended range
225, 226 & 228: ±5% of reading outside high recommended range
T)
High performance 225 Toroidal &
226 Conductivity sensors
Repeatability: ±0.25% ±5 µS/cm after
10mS/cm
100mS/cm
500mS/cm to 2000mS/cm
1500mS/cm to 2000mS/cm
Instruction Manual
LIQ-MAN-1066
1000mS/cm
2000mS/cm
Specifications

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents