Section 10: haRT Communication
April 2017
1066 ph/ORp
continued
ph_display_resolution
Changes the resolution of the displayed pH value.
ph_sensor_isopotential
The pH at which the millivolt output of the pH sensor remains
constant with temperature changes; virtually always 7.00 pH.
refz_high_fault_setpoint
The high setpoint that triggers a reference electrode
mpedance alert.
glassz_high_fault_setpoint
The high setpoint that triggers a glass pH electrode impedance
alert.
diagnostics_switch
Turns the glass and reference electrode impedance
measurements on and off.
glassz_temperature_correction
Turns temperature correction of the glass impedance
measurement on and off.
glassz_measurement_type
Toggles between a basic glass impedance measurement, and
an advanced impedance measurement that is more accurate.
calculated_offset
The preliminary zero offset calculated by a pH
calibration before it is accepted by the transmitter.
calculated_slope
The preliminary slope calculated by a pH calibration before it is
accepted by the transmitter.
110
Type
Enumeration / Range
ENUM
0.01 pH
0.1 pH
FLOAT
0.00 -- 14.00 pH
FLOAT
0.0 -- 9,999.0 kohm
FLOAT
0.0 -- 9,999.0 Mohm
ENUM
Off
On
ENUM
Off
On
ENUM
Advanced
Basic
FLOAT
FLOAT
Instruction Manual
LIQ-MAN-1066
Default
Read/Write
R / W
0.01 pH
R / W
7.00 pH
R / W
500.0 kohm
R / W
1,500 Mohm
R / W
Off
R / W
Off
R / W
Advanced
R / W
R / W
HART Communications