Section 4: Wiring
April 2017
4.3
power Supply Wiring For 1066-FF

Power Supply Wiring

4.3.1
Run the power/signal wiring through the opening nearest TB2. Use shielded cable and ground the
shield at the power supply. To ground the transmitter, attach the shield to TB2-3.
Note: For optimum EMI/RFI immunity, the power supply/output cable should be shielded and
enclosed in an earth-grounded metal conduit. Do not run power supply/signal wiring in the same
conduit or cable tray with loop power lines. Keep power supply/signal wiring at least 6 ft (2 m)
away from heavy electrical equipment.
F
OuNDaTION
Figure 4-4
water. The figure also shows three ways in which Fieldbus communication can be used to read
process variables and configure the transmitter.
FIGuRE 4-4. Configuring 1066p Transmitter with F
FIGuRE 4-5. Typical Fieldbus Network Electrical Wiring Configuration
20
Fieldbus
shows a 1066PFF being used to measure and control pH and chlorine levels in drinking