Calibrating Temperature; Calibration - Emerson Rosemount 1066 Instruction Manual

Smart-enabled, 2-wire transmitter
Hide thumbs Also See for Rosemount 1066:
Table of Contents

Advertisement

Instruction Manual
LIQ-MAN-1066
9.9

Calibrating Temperature

Most liquid analytical measurements require temperature compensation (except ORP and Redox).
The 1066 performs temperature compensation automatically by applying internal temperature
correction algorithms. Temperature correction can also be turned off. If temperature correction is
off, the 1066 uses the manual temperature entered by the user in all temperature correction
calculations.
To calibrate temperature, access the Calibration screen by pressing ENTER from the main screen,
select Temperature and press ENTER.
The following calibration routine is covered:
1. Temperature with manual temperature entry
To calibrate temperature:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Temperature. Press ENTER.
The adjacent screen will appear.
The following sub-section provides you with the initial display screen that appears for temperature
calibration. Use the flow diagram for Temp calibration at the end of Sec. 7 to complete calibration.
9.9.1

Calibration

The adjacent screen will appear during Temperature Cal.
If the sensor Temperature offset is greater than 5 °C from the default
value, the following screen will appear:
You may continue by selecting Yes or suspend this operation by selecting
No.
If the Temp Cal is successful, the screen will return to the Cal Menu.
Note: To select automatic or manual temp compensation or to program temperature units as °C
or °F, refer to
Calibration
Section 7.3
– Programming Temperature in this manual.
Section 9: Calibration
April 2017
1.234 µS/cm
25.0 ºC
SN Calibrate
+0 25.0 °C
1.234 µS/cm
25. 0 ºC
SN Calibrate
Cal in progress.
Please wait.
1.234 µS/cm
25.0 ºC
SN Temp Offset > 5 °C
Continue?
Yes
No
85