Instruction Manual
LIQ-MAN-1066
Solution Temperature Correction
8.2.4
The display screen for selecting the Solution temperature correction
algorithm is shown. The default value is displayed in bold type. Refer
to the pH/ORP Programming flow diagram to complete this function.
8.2.5
Temperature Coefficient
The display screen for entering the custom solution temperature
coefficient is shown. The default value is displayed in bold type. Refer
to the pH/ORP Programming flow diagram to complete this function.
8.2.6
Resolution
The display screen for selecting 0.01pH or 0.1pH for pH display
resolution is shown. The default value is displayed in bold type. Refer
to the pH/ORP Programming flow diagram to complete this function.
8.2.7
Filter
The display screen for entering the input filter value in seconds is
shown. The default value is displayed in bold type. Refer to the
pH/ORP Programming flow diagram to complete this function.
8.2.8
Reference Impedance
The display screen for selecting Low or High Reference impedance is
shown. The default value is displayed in bold type. Refer to the
pH/ORP Programming flow diagram to complete this function.
8.3
ORp Measurement programming
The section describes how to configure the 1066 transmitter for ORP measurements. The
following programming and configuration functions are covered:
1. Measurement type: pH Select pH, ORP, Redox.
2. Preamp location: Transmitter Identify preamp location
3. Filter: 4 sec Override the default input filter, enter 0-999 seconds
4. Reference Z: Low Select low or high reference impedance
5. Sensor wiring scheme: Normal or Reference to Ground
Programming Measurements
Section 8: programming Measurements
1.234 µS/cm
SN Sol'n Temp Corr.
Ultra Pure Water
Off
High pH
Custom
1.234µS/cm
SN Sol'n Temp Coeff.
- 0.032pH/ºC
1.234 µS/cm
SN Resolution
0.1pH
0.01pH
1.234 µS/cm
SN Input filter
04 sec
1.234 µS/cm
SN Reference Z
High
Low
April 2017
25.0 ºC
25.0ºC
25.0 ºC
25.0 ºC
25.0 ºC
45