Section 3: Installation

3.1

unpacking and inspection

Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions.
Save the box. If there is no apparent damage, unpack the container. Be sure all items shown on the
packing list are present. If items are missing, notify Rosemount immediately.
Installation – General Information
3.2
1. Although the transmitter is suitable for outdoor use, installation is direct sunlight or in areas
of extreme temperatures is not recommended unless a sunshield is used.
2. Install the transmitter in an area where vibration and electromagnetic and radio frequency
interference are minimized or absent.
3. Keep the transmitter and sensor wiring at least one foot from high voltage conductors. Be sure
there is easy access to the transmitter.
4. The transmitter is suitable for panel, pipe, or surface mounting.
5. The transmitter case has six 1/2-inch (PG13.5) conduit openings. Use separate conduit
openings for the power/output cable, the sensor cable, and the other the sensor cable as
needed (pH input for free chlorine with continuous pH correction).
6. Use weathertight cable glands to keep moisture out to the transmitter. If conduit is used, plug
and seal the connections at the transmitter housing to prevent moisture from getting inside
the instrument.

preparing Conduit Openings

3.3
There are six conduit openings in all configurations of 1066.
Conduit openings accept 1/2-inch conduit fittings or PG13.5 cable glands. To keep the case
watertight, block unused openings with Type 4X or IP66 conduit plugs.
To maintain ingress protection for outdoor use, seal unused conduit holes with suitable conduit
plugs.
NOTE: Use watertight fittings and hubs that comply with your requirements. Connect the conduit
hub to the conduit before attaching the fitting to the transmitter.
Electrical installation must be in accordance with the National Electrical Code (aNSI/NFpa-70) and/or
any other applicable national or local codes.