1066Oz/Cl/DO
polar_voltage
The polarization voltage used by the transmitter; automatically
set by sensor selection.
temp_coeff
The temperature coefficient used to compensate temperature;
automatically set by sensor selection.
amp_sensor_sensitivity
The response of the sensor to changes in concentration, deter-
mined by calibration.
amp_sensor_zero_current
The current output of the sensor when the concentration is 0.
It is determined by a zero calibration.
1066Oz
chlorine_type
The resolution of the displayed ozone measurement.
1066Cl
chlorine_type
The type of chlorine being measured.
chlorine_resolution
The resolution of the displayed chlorine measurement.
ph_correction_mode
Turns automatic pH compensation of free chlorine measure-
ments on and off.
manual_ph
Provides a constant pH value, if manual pH compensation of a
free chlorine measurement is used.
HART Communications
Section 10: haRT Communication
Type
Enumeration / Range
FLOAT
FLOAT
FLOAT
0.1 -- 1,000,000.0
nA/ppm
FLOAT
-999.9 -- 999.9 nA
Type
Enumeration / Range
ENUM
0.001
0.01
Type
Enumeration / Range
ENUM
Free Chlorine
Total Chlorine
Chloramine
ENUM
0.001 ppm
0.01 ppm
ENUM
On
Off
FLOAT
0.00 -- 14.00 pH
April 2017
Default
Read/Write
R / W
0
R / W
0
R / W
2,500 nA/ppm
R / W
0.0 nA
Default
Read/Write
R / W
0.001
Default
Read/Write
R / W
Free Chlorine
R / W
0.001 ppm
R / W
Off
R / W
7.00 pH
113