Carrying Out The Fresh Air Calibration - Dräger X-AM 5000 Technical Manual

Multi-gas monitor
Hide thumbs Also See for X-AM 5000:
Table of Contents

Advertisement

Maintenance

Carrying out the fresh air calibration

To improve the zero point accuracy, you can carry out a fresh air calibration.
If none of the sensors fitted permits calibration with fresh air (e.g. only O
CO
), fresh air calibration is not offered as a menu function.
2
Calibrate the device to fresh air, free of measured gases or other interfering gases.
Not all sensors are included in the fresh air calibration
warmed up or which are faulty prevent a calibration.
In the case of sensors which are in the warm-up phase, the message » 159 « is dis-
played with the special symbol »
In the case of a sensor or device error, the message » 109 « is displayed with the
special symbol »
The message is cleared after 5 seconds and the function is available again in the
menu
During the fresh air calibration, the zero point of all sensors (with the exception of
DrägerSensor XXS O
In the case of DrägerSensor XXS O
Switch on the device.
Depending on device configuration:
Call the quick menu and select the
Fresh Air Calibration function,
page 16.
or
Call the calibration menu and select
the Fresh Air Calibration function,
page 17.
The current gas concentration values
flash.
When the measured values have sta-
bilized:
Press the
key to perform the fresh
OK
air calibration.
1)
Fresh air calibration / zero point adjustment is not supported by the DrägerSensor XXS O
point calibration / adjustment of these sensors can be conducted using the Dräger CC-Vision PC soft-
ware. To do so, a suitable zero gas that is free of ozone (e.g. N
38
NOTICE
« (for warning message).
« (for a fault message).
) is set to 0.
2
, the display is set to 20.9 vol. %.
2
, only IR-
3
1)
. Sensors which have not
ch
4
%LEL
O
2
Vol%
CO
ppm
H
S
2
ppm
NH
3
ppm
) should be used.
2
. A zero
3