Gebrauchsanweisung Käyttöohjeet Návod na použití Instructions for Use Bruksanvisning Notice d’utilisation Bruksanvisning Instruc iuni de utilizare Instrucciones de uso Instrukcja obs ugi Használati útmutató Instruções de uso Istruzioni per l'uso Upute za uporabu Kullanma talimatlar Gebruiksaanwijzing Navodilo za uporabo Brugsanvisning Návod na použitie ®...
Safe coupling with electrical devices NOTICE Electrical connections to devices which are not listed in these Instructions Indicates additional information on how to use the product. for Use should only be made following consultation with the respective manufacturers or an expert. Dräger X-am 2500...
Display 14 Special symbols or dusts of groups A, B, C, D or E, F, G, and temperature class T3 or T4 (depending on the batteries and rechargeable battery) may 15 Tool for sensor change be present. Dräger X-am 2500...
Dräger. Different settings can be selected to meet customer requirements on delivery. The current setting can be checked and changed with the Dräger CC Vision software. Only when activated in the instrument configuration. Delivery condition: not activated. Dräger X-am 2500...
In the event of an impact load when using a CatEx sensor in the Dräger X-am 2500 that causes the fresh air display to In case of measurements in oxygen-deficient atmosphere deviate from zero, the zero point and sensitivity must be (<8 vol.
A1 = lack of oxygen 4.6.5 Battery main alarm A2 = excess oxygen Intermittent alarm: After leaving the area, when the concentration has dropped below the alarm threshold: Flashing special symbol on the right side of the display: Dräger X-am 2500...
Press the [+]-key when the instrument is turned off. The name of the gas, measuring unit, and measuring range limit value are displayed for all channels. Press the [+]-key again to exit the Info Off Mode (or via timeout). Dräger X-am 2500...
For battery holder (order no. 83 22 237): replace alkaline batteries or three complete charge/discharge cycles. Never store the instrument for respectively NiMH rechargeable batteries. Ensure correct polarity. extended periods without being connected to a power source (maximum of 2 months) because the internal buffer battery will drain. Dräger X-am 2500...
Page 10
Have the instrument calibrated by the service personnel. selected sensor is set to the value of the test gas used. Use a standard test gas. Upon application of the Dräger mixed gas (order no. 68 11 130) the displays should be within this range. Dräger X-am 2500...
Page 11
The measurement value flashes. For the adjustment of the ex-channel to the measuring gas methane, the display of the instrument must be set to a value of 5 % (relative) higher than the test gas concentration used. Dräger X-am 2500...
Collect batteries and rechargeable batteries as specified by Abrasive cleaning tools (brushes etc.), cleaning agents and the applicable regulations and dispose of at battery collection cleaning solvents can destroy the dust and water filters. centres. Carefully dry the instrument with a cloth. Dräger X-am 2500...
220 to 250 g The Technical Handbook and Instructions for Use/data sheets for the utilized sensors and the PC software CC-Vision for Dräger X-am 2500 can be downloaded on the product page of the X-am 2500 at the following Internet address: www.draeger.com Not subject to BVS10 ATEX E 080X and PFG 10 G 001X performance approval.
Page 14
EN 50271 The Technical Handbook and Instructions for Use/data sheets for the utilized sensors and the PC software CC-Vision for Dräger X-am 2500 can be downloaded on the product page of the X-am 2500 at the following Internet address: www.draeger.com For decreasing concentrations, the adjustment time for nonane is about 50 seconds.
Page 15
Effect of sensor poisons – – – – – – Hydrogen sulphide H S, 10 ppm Hydrogen sulphides, heavy metals, silicone-, sulphur- and polymerization-capable materials ≤±2 % of the measuring ≤±2 % of the measuring Linearity error value value Dräger X-am 2500...
D = 2012, E = 2013, F = 2014, H = 2015, J = 2016, K = 2017, L = 2018 etc. Example: Serial No. AREH-0054: the third letter is E, which means that the Not subject to BVS10 ATEX E 080X and PFG 10 G 001X performance ap- unit was manufactured 2013. proval. Dräger X-am 2500...
Use only battery packs type ABT 01xx, HBT 00xx or HBT 01xx. See marking on battery pack for approved batteries and related Temperature Code. Not tested in oxygen enriched atmospheres (>21 % O High off scale readings may indicate an explosive concentra- tion. Dräger X-am 2500...