Stel / Twa Exposure Alarm; Battery Pre-Alarm; Battery Main Alarm; Device Alarm - Dräger X-AM 5000 Technical Manual

Multi-gas monitor
Hide thumbs Also See for X-AM 5000:
Table of Contents

Advertisement

Identifying Alarms

STEL / TWA exposure alarm

The alarm is indicated by an intermittent alarm message:
Display » A2 « and »
value alternating:
Leave the area immediately. After this alarm, the deployment of per-
sonnel is subject to the relevant national regulations.
STEL and TWA alarms cannot be acknowledged or canceled.
Switch off the device. The values for the exposure evaluation are deleted after the
device is switched on again.

Battery pre-alarm

The alarm is indicated by an intermittent alarm message:
Flashing special symbol »
Acknowledging the pre-alarm:
Press the
key. Only the audible alarm and the vibration alarm are switched off.
OK
The battery still lasts approx. 20 minutes after the first battery pre-alarm.

Battery main alarm

The alarm is indicated by an intermittent alarm message:
Flashing special symbol »
The battery main alarm cannot be acknowledged or canceled:
The device is automatically switched off again after 10 seconds.
Before the device is switched off, the visual, audible and vibration alarms are activa-
ted for a short time.

Device alarm

The alarm is indicated by an intermittent alarm message:
Special symbol »
The device or one or several sensor channels are not ready for operation.
For remedies, refer to page 24 to page 30.
If necessary, commission the Dräger Service Center to eliminate the error.
20
« (TWA) or »
CAUTION
« on the right side of the display:
« on the right side of the display:
« on the right side of the display:
« (STEL) and measured