Quick menu "Displaying and
deleting peak values"
After the function has been selected, the
current peak values are displayed; the
peak values special symbol appears in
the display at the same time.
•
The peak values can be deleted by
pressing the
OK
The adjacent display appears.
•
Press the
key to end the function.
OK
Calling the Calibration Menu
−
The calibration menu can only be accessed by entering a password.
Password on delivery: » 001 «
−
The default password on delivery can be changed using the PC software
Dräger CC Vision.
•
In measuring mode, press the
−
The function for entering the password is selected.
−
The »
« special symbol (for the "enter password" function) is displayed.
key for 5 sec.
key for at least 4 seconds.