Dräger X-AM 5000 Technical Manual page 30

Multi-gas monitor
Hide thumbs Also See for X-AM 5000:
Table of Contents

Advertisement

Faults, Cause and Remedy
Special symbol »
and displayed numerical
code:
505
506
507
524
525
526
601
602
603
604
605
606
607
624
626
30
«
Cause
Error during the function test
with gas (bump test) of Drä-
gerSensor XXS EC3
Failed filter test.
Failed rise time test.
Device incorrectly configured
by Dräger CC-Vision.
The span calibration for the
compensation channel is not
valid
Error during warm-up accele-
ration Dräger Sensor XXS
EC3
The zero point calibration of
DrägerSensor XXS EC4 is not
valid
The span calibration of
DrägerSensor XXS EC4 is not
valid
The measured value of Drä-
gerSensor XXS EC4 is in the
negative range
DrägerSensor XXS EC4 is not
inserted or defective
Error during the function test
with gas (bump test) of Drä-
gerSensor XXS EC4
Failed filter test.
Failed rise time test.
Device incorrectly configured
by Dräger CC-Vision.
Error during warm-up accele-
ration Dräger Sensor XXS
EC4
Remedy
Repeat function test. Calibrate
or replace DrägerSensor XXS
EC3, if necessary page 49.
Repeat filter test with X-dock.
Repeat rise time test with X-
dock.
Change sensor for applicable
channel with Dräger CC-Vision.
Carry out span calibration for
compensation electrode.
Disconnect and reconnect
power pack or replace the sen-
sor. Sensor must not be loaded
with gas within the first 5 minu-
tes.
Carry out fresh air calibration,
page 38.
Carry out span calibration,
page 42.
Carry out fresh air calibration,
page 38.
Check DrägerSensor XXS EC4,
page 49.
Repeat function test. Calibrate
or replace DrägerSensor XXS
EC 4, if necessary page 49.
Repeat filter test with X-dock.
Repeat rise time test with X-
dock.
Change sensor for applicable
channel with Dräger CC-Vision.
Disconnect and reconnect
power pack or replace the sen-
sor. Sensor must not be loaded
with gas within the first 5 minu-
tes.