Calibration
3. Attach the calibration adapter to the sensor.
1
Fig. 1
Test gas cylinder (1), test gas adapter (2)
6.3
Carrying out the zero calibration
The zero calibration can be accessed via the "Calibration" menu item
in the live overview.
The zero calibration can be interrupted at any time. To interrupt the
calibration, select Back.
70
On this page, the following information about the inserted sensor will
be shown (see figure):
– The date of the last calibration
– The part number
– The date of manufacture
Not every sensor supports a zero calibration.
2
Technical manual
|
Dräger X-node