Fault Messages - Dräger X-AM 5000 Technical Manual

Multi-gas monitor
Hide thumbs Also See for X-AM 5000:
Table of Contents

Advertisement

Fault messages

Special symbol »
and displayed numerical
code:
102
103
104
105
106
107
108
109
111
112
113
114
115
116
201
202
«
Cause
The customer's service life
counter has elapsed
The device is defective
Check sum error program
code
The bump test interval has
elapsed
The calibration interval has
elapsed (at least 1 calibration
interval has elapsed)
Bump test error (at least
1 channel has a bump test
error)
The device is defective
The menu function cannot be
carried out because of an
error.
Failed alarm element test:
alarm light.
Failed alarm element test:
alarm horn.
Failed alarm element test:
Vibration motor.
Failed visual inspection.
Device is disabled by
X-dock.
Failed software update.
The zero point calibration of
DrägerSensor CatEx 125 PR
is not valid
The span calibration of Drä-
gerSensor CatEx 125 PR is
not valid
Faults, Cause and Remedy
Remedy
Reset the service life counter
using Dräger CC Vision.
The device must be repaired by
DrägerService.
The device must be repaired by
DrägerService.
Carry out bump test, page 35.
Carry out span calibration,
page 40 or page 42.
Carry out bump test, page 35 or
carry out span calibration,
page 40 or page 42.
The device must be repaired by
DrägerService.
Determine the error code via the
info menu and switch it off, if
necessary.
Repeat alarm element test
with Dräger X-dock.
Repeat alarm element test
with X-dock.
Repeat alarm element test
with X-dock.
Repeat visual inspection with
X-dock.
Activate device with X-dock.
The device must be repaired by
DrägerService.
Carry out fresh air calibration,
page 38.
Carry out span calibration,
page 40 or page 42.
27