Span calibration for CatEx

If the measuring range end value < = 100%LEL, the adjustment for the heat tinting
is offered.
Display in the case of channel selection:
Open the test gas cylinder valve to let
test gas flow over the sensor.
Press the
key to start the calibrati-
OK
on for the heat tinting or press the
-key to select the next sensor.
If the displayed measurement value is
stable:
Press the
key to perform the
OK
calibration.
The display containing the current
gas concentration changes with the
display » OK «.
Press the
key or wait for approx. 5
OK
seconds to end the calibration of this
measuring channel.
The next measuring channel is
offered for calibration.
After the calibration of the last
measuring channel, the device
changes to measuring mode.
Close the test gas cylinder valve and remove the device from the calibration cradle.