Open the test gas cylinder valve to let test gas flow over the sensor.
The currently displayed measured values start to flash.
The flashing stops after a static measured value has been reached.
The calibration is now carried out automatically.
The displayed measured values change to the values according to the gas supplied.
The automatic stability monitoring can be overridden by pressing the OK key. A ca-
libration then takes place immediately. If it is detected that no test gas has been ap-
plied, the 1-button calibration will be aborted. The channels will then indicate » n/
a «. If only one sensor is taking part in the 1-button calibration, a calibration will be
performed in each case when the OK key is pressed.
When the calibration is completed
and the displayed measured values
have stabilized:
The display containing the current
gas concentration changes with the
display » OK «.
Press the
key or wait for 5 se-
OK
conds to quit the calibration.
The device changes to the measuring
mode
Close the test gas cylinder valve and
remove the device from the calibrati-
on cradle.
If a fault occurred during the 1-button
calibration.
The fault message »
and »
« is displayed for the res-
pective sensor instead of the measu-
red value.
In this case, repeat the 1-button calib-
ration or carry out a single gas calib-
ration, refer to page 42.
If necessary, replace the sensor,
page 49.
« appears
Maintenance
ch
4
%LEL
O
2
Vol%
CO
ppm
H
S
2
ppm
ch
4
%LEL
O
2
Vol%
CO
ppm
H
S
2
ppm
41