Faults, Cause and Remedy
Special symbol »
and displayed numerical
code:
456
457
551
552
553
554
555
556
557
575
576
651
652
653
654
655
656
657
26
«
Cause
The calibration interval for
DrägerSensor XXS EC2 has
elapsed
Alarm setpoint A2 setting is
greater than 60 %LEL
DrägerSensor XXS EC3 in
the warm-up phase
DrägerSensor XXS EC3 in
the warm-up phase
EC3 concentration has drifted
into the negative range
The temperature is too high
The temperature is too low
The calibration interval for
DrägerSensor XXS EC3 has
elapsed
Alarm setpoint A2 setting is
greater than 60 %LEL
The calibration interval for the
compensation electrode has
elapsed
Calibration required because
of overgassing.
DrägerSensor XXS EC 4 in
the warm-up phase
DrägerSensor XXS EC 4 in
the warm-up phase
EC 4 concentration has drifted
into the negative range
The temperature is too high
The temperature is too low
The calibration interval for
DrägerSensor XXS EC 4 has
elapsed
Alarm setpoint A2 setting is
greater than 60 %LEL
Remedy
Carry out span calibration for
DrägerSensor XXS EC 3,
page 42.
Set alarm setpoint to less than
60 %LEL.
Wait until warm-up time is com-
plete.
Wait until warm-up time is
complete.
Carry out fresh air calibration,
page 38.
Operate the device within the
allowed temperature range.
Operate the device within the
allowed temperature range.
Carry out span calibration for
DrägerSensor XXS EC 3,
page 42.
Set alarm setpoint to less than
60 %LEL.
Carry out span calibration for
compensation electrode.
Perform a span calibration for
the compensation electrode.
Wait until warm-up time is
complete.
Wait until warm-up time is
complete.
Carry out fresh air calibration,
page 38.
Operate the device within the
allowed temperature range.
Operate the device within the
allowed temperature range.
Carry out span calibration for
DrägerSensor XXS EC 4,
page 42.
Set alarm setpoint to less than
60 %LEL.