Care

Start of the warm-up acceleration.
The remaining time is displayed and
decremented.
WARNING
This function requires that the device is
in a gas-free environment. A warning
screen is displayed and the remaining
time during which no gas may be
applied is shown It is essential to
ensure that the sensor is in the fresh air
during this time (for approx. 5 minutes)!
Any gas applied during this period could
later result in faulty alarms or sup-
pressed alarms!
If the procedure is unsuccessful, e.g. as a result of a defective sensor, this will be indi-
cated by error code » X07 « on the channel for the affected sensor (X). In the event of
a fault, the device will also switch to the unaccelerated mode.
Dräger recommends that the device not be switched off and on again for a period of
two hours when the accelerated warm-up is in use.
Care
The device does not need any special care.
Dirt and deposits can be removed from the device by washing it with cold water. A
sponge can be used for wiping if necessary.
Abrasive cleaning implements (brushes etc.), cleaning agents and cleaning solvents
can destroy the dust and water filters.
Carefully dab dry the device using a cloth.