Maintenance; Maintenance Intervals - Dräger X-AM 5000 Technical Manual

Multi-gas monitor
Hide thumbs Also See for X-AM 5000:
Table of Contents

Advertisement

Maintenance

Maintenance intervals

The device should be inspected and maintained by suitably qualified persons annually
(consult: EN 60079-29-2 – Gas measuring device - Selection, installation, use and
maintenance of apparatus for the measurement of combustible gases or oxygen,
EN 45544-4 – Electrical apparatus used for the direct detection and direct concentra-
tion measurement of toxic gases and vapors - Part 4: Guide for selection, installation,
use and maintenance and national regulations).
Recommended calibration interval for measuring channels Ex, O
months.
Calibration interval of other gases: refer to the Instructions for Use of the respective
DrägerSensors.
Depending on device configuration:
Replace the alkaline batteries or charge the battery – refer to page 45 to page 46 –
after each use, at the latest after the battery alarm has been triggered or after 2
weeks.
Calibrating the device – page 37.
In regular intervals, according to the sensors used and the operating conditions. For
sensor-specific calibration data, refer to the Instructions for Use/data sheets of the
1)
sensors used
.
Before you carry out safety-related relevant measurements, the zero point and sen-
sitivity of the devices should be tested in accordance with national regulations.
Inspection by suitably qualified persons – every year.
The inspection intervals must be established in each individual case and shortened
if necessary, depending on technical safety considerations, engineering conditions
and the technical requirements of the equipment.
We recommend that a service agreement be concluded with Dräger and that repairs
also be carried out by them.
Replace the sensors, page 49 – if necessary, when it is not possible to calibrate the
sensors anymore.
1)
Instructions for Use/data sheets for DrägerSensors can be downloaded on the product page of the X-am
5000 at the following Internet address: www.draeger.com. See also the enclosed Instructions for Use and
data sheets for the sensors used.
Maintenance
, H
S and CO: 6
2
2
31