Indicates a potentially hazardous situation which, if not avoided, could result in physical injury, or damage to the product or environment. It may also be used to alert against unsafe practices. NOTICE Indicates additional information on how to use the product. Dräger X-am 2500...
Function for password input is active, see section 4.4 on page 17 Fastening clip Nameplate Battery / rechargeable battery 100 % full Charging contacts Battery / rechargeable battery 2/3 full Power pack Battery / rechargeable battery 1/3 full Serial no. Battery / rechargeable battery empty Dräger X-am 2500...
Do not stick anything on the name plate on the gas detector. with the installed Dräger sensors. The technical approvals are valid for the X-am 2500 gas detection instrument and the calibration cradle. The explosion- Areas subject to explosion hazards, classified by zones...
2 months) because the internal buffer battery will drain. modules, e. g., by metallic objects that have fallen in, does not result in damage to the charging station. It should, however, be avoided due to possible heating hazards and incorrect displays on the charger module. Dräger X-am 2500...
The charging process is carried out analogous to charging manufacturer's specifications. Ambient temperature with the multiple charging station. during the charging process: 0 to +40 °C. 1) Not part of the measurement performance tests BVS10 ATEX E 080X and PFG 10 G 001X. Dräger X-am 2500...
(e.g. not yet ready for calibration) are active. explosive concentration where applicable. 2. Press the [OK] key to cancel the display of the activation sequence. 1) Only when activated in the instrument configuration. Delivery condition: not activated. Dräger X-am 2500...
1) Different settings can be selected to meet customer requirements on delivery. The current setting can be checked and changed with the Dräger CC Vision software. A version of the CC-Vision software that can be used for Dräger X-am 2500 is available for download from the product page for the X-am 2500 at the following web address: www.draeger.com 2) With O , A1 is the lower alarm threshold: an alarm is triggered if the value is too low.
Time (time on the PC) The CC-Vision PC software can be used to select or disable Warning after expiry of Yes/No the capture ranges for the measuring mode. calibration interval Error after expiry of calibration Yes/No interval Dräger X-am 2500...
Vol%, %UEG, %LEL, %LIE, ppm, mbar, ppb, mg/m Gas name: “Ex” Yes/No (CatEx sensor only) Automatic fresh air calibration Yes/No in the charging cradle (CatEx sensor only) 1) Only evaluated if the sensor is provided for the purpose. Dräger X-am 2500...
If the set bump test concentration is not reached within the set time, an error is issued. Upon application of the Dräger mixed gas (order no. 68 11 130) the displays should be within this range. Dräger X-am 2500...
NOTICE 1. Prepare the Bump Test Station according to the instructions. After the bump test, the display shows a printer icon even if there is no printer connected to the bump gas station. Dräger X-am 2500...
The instrument is not ready for operation. For corrective measures, see“Replacing the sensors” on page 22 to page 25. Contact maintenance or Draeger Service to rectify the problem. 1) Only when activated in the instrument configuration. Delivery status: not activated. Dräger X-am 2500...
00433366_en.eps 1. The peak values can be deleted by pressing the key for 5 sec. and the %LEL adjacent display appears, for example. Vol% 2. Press the key to end the function. 00533366_en.eps Dräger X-am 2500...
Vol% Warming-up time: see instructions for use / data sheets for out the fresh air calibration. the Dräger sensors installed (product page for X-am 2500 at www.draeger.com). Adjustment interval: Observe the relevant specifications in the Instructions for ...
5.4.1 on page 20. 2. Vent the test gas into a fume cupboard or into the open Replace sensor air (with a hose connected 00833366.eps necessary, page 22. to the second connector of the calibration cradle). 01833366_en.eps Dräger X-am 2500...
6. Open the test gas cylinder valve to let test gas flow over the sensor. The displayed, flashing measurement value changes to the value according to the supplied test gas. 7. Press the -key to select the next sensor. Dräger X-am 2500...
Leak test: pump battery: 2/3 Block suction inlet Remaining charge of Leak test: pump battery: 1/3 Release suction inlet Pump battery discharged Commissioning and performing the measurement: See instructions for use of the Dräger X-am Pump. Dräger X-am 2500...
Conduct the fresh air calibration page 18. and then: Commissioning and performing the measurement: Calibrating sensitivity: See instructions for use of the accessories used. either perform 1-button calibration, page 19 run sensitivity calibration, see page 20. Dräger X-am 2500...
DrägerSensor XXS EC1 is warming up Wait until warm-up time is complete. DrägerSensor XXS EC1 is warming up Wait until warm-up time is complete. EC1 concentration has drifted into the Conduct the fresh air calibration page 18. negative range Dräger X-am 2500...
Page 24
The calibration interval for DrägerSensor Run sensitivity calibration for DrägerSensor XXS EC 4 has elapsed XXS EC 4, page 20. Alarm threshold A2 is set to greater than Set alarm threshold to less than 60 %LEL. 60 %LEL Dräger X-am 2500...
19 and/or page 20. conduction Instrument incorrectly configured by Dräger Change sensor for applicable channel with Dräger CC-Vision. CC-Vision. No valid zero point calibration of the Dräger Conduct the fresh air calibration page 18. Sensor XXS EC1 Dräger X-am 2500...
Page 26
Carry out span calibration for compensation compensation channel electrode. Error during warm-up acceleration Dräger Disconnect and reconnect power pack or replace Sensor XXS EC3 the sensor. Sensor must not be loaded with gas within the first 5 minutes. Dräger X-am 2500...
Page 27
Change sensor for applicable channel with Dräger CC-Vision. CC-Vision. Error during warm-up acceleration Dräger Disconnect and reconnect power pack or replace Sensor XXS EC4 the sensor. Sensor must not be loaded with gas within the first 5 minutes. Dräger X-am 2500...
1) Instructions for use/data sheets for the Dräger sensors can be downloaded from the product page for the X-am 2500 on the following website: www.draeger.com. See also the enclosed instructions for use and data sheets for the sensors used.
Technical data Technical data 12.1 X-am 2500 Ambient conditions: during operation and storage Temperature class T4 (–20 to +50 °C): NiMH power packs type: HBT 0000, HBT 0100 Power pack type: ABT 0100 with alkaline single cell type: Duracell Procell MN 1500 Duracell Plus Power MN 1500 Temperature class T3 (–20 to +40 °C):...
With default alarm thresholds that can be Pump) adjusted specifically for each country. USB cable (for Dräger X-am Pump) 83 27 108 Dräger X-am 2500 standard instrument: Rubber ball pump 68 01 933 Dräger X-am 2500 Ex 83 23 910...
3) This accessory is optimised for the Dräger X-am Pump (market launch in 2017) (for hoses with 3 mm inner diameter). 4) 3 mm tube connection set (order no. 83 27 641) required. 5) Expected sensor life time: O and CO >5 years, CatEx > 3 years. Dräger X-am 2500...