Calibrating The Sensitivity For An Individual Measuring Channel - Dräger X-AM 5000 Technical Manual

Multi-gas monitor
Hide thumbs Also See for X-AM 5000:
Table of Contents

Advertisement

Maintenance

Calibrating the sensitivity for an individual measuring channel

The span calibration can be carried out specifically for individual sensors.
In the case of the span calibration, the sensitivity of the selected sensor is set to the
value of the test gas used.
Use a standard test gas.
Allowed test gas concentration:
Ex:
40 to 100 %LEL
O
10 to 25 vol. %
2
CO: 20 to 999 ppm
H
S: 5 to 99 ppm
2
Test gas concentration of other gases:
refer to the Instructions for Use of the
respective DrägerSensors.
Connect the test gas cylinder with the
calibration cradle.
Vent the test gas into a fume cup-
board or into the open air (with a hose
connected to the second connector of
the calibration cradle).
CAUTION
Do not inhale the test gas. Risk to
health!
Observe the hazard warnings of the
relevant Safety Data Sheets.
Switch on the device and insert it into the calibration cradle.
Press and hold the [ + ] key for 5 seconds to call the calibration menu, enter the
password and select the single gas calibration function, page 17.
Press the
key to start the channel
OK
selection.
The display flashes the gas of the first
measuring channel, e.g., » Ex - %LEL
«.
Press the
key to carry out the cali-
OK
bration of this measuring channel, or
Use the
key to select another
measuring channel (O
- ppm or CO - ppm).
42
- vol. %, H
S
2
2
0.5 L/min
X-am 5000
CH
4
%LEL
O
2
Vol%
CO
ppm
H
S
2
ppm
0
ch
4
%LEL
CO
2
ppm
O
2
Vol%
CO
ppm
H
S
2
ppm
NH
3
ppm