Activating the Device

Standard device configuration

Function test with gas (bump test)
1)
in Quick Menu
Fresh air calibration in Quick Menu
Life sign - optical only
1)
Switch off
1)
LEL factor
(ch
4
1)
Averaging time
1)
Different settings can be selected to meet customer requirements on delivery. The current setting can
be checked and changed with the Dräger CC Vision software. A version of Dräger CC-Vision suitable
for the Dräger X-am 5000/2000 can be downloaded on the product page of the X-am 5000: www.drae-
ger.com.
Changing the standard configuration: See "Configuring the Device" on page 22.
After a basic initialization has been carried out with the PC software Dräger CC
Vision, individual alarm settings may have been changed.
Activating the Device
Before using the device for the first time, insert the supplied batteries or a charged
NiMH power pack T4 (order no. 83 18 704), refer to Replacing the Batteries, page 45.
Charge the rechargeable batteries if necessary, page 46.
The Dräger X-am 5000 is ready for operation.
12
1)
1)
)
Quick bump test
on
on
allowed
4.4 vol. %
(4.4 vol. % corresponds to 100 %LEL)
15 minutes for STEL
8 hours for TWA