Maintenance
If a fault occurred during the span
calibration.
The fault message »
and »
« is displayed for the
sensor instead of the measured
value.
In this case, repeat the calibration.
If necessary, replace the sensor,
page 49.
If the gas CH
(measurement range
4
up to 100 vol. %) is selected on the
CatEx channel, the heat conductivity
is calibrated, display with
channelselection:
Press the
key to perform the
OK
calibration for heat conductivity.
Press the
key to select the next
sensor.
Press the
key to perform the
OK
calibration of the selected measuring
channel.
The test gas concentration is
displayed.
Press the
key to confirm the test
OK
gas concentration or use the [ + ] key
to change the test gas concentration
and complete the process by
pressing the
OK
The measurement value flashes.
Open the test gas cylinder valve to let
test gas flow over the sensor.
The displayed, flashing measurement value changes to the value according to the
supplied test gas.
Press the
key to select the next sensor.
44
« appears
key.
ch
4
%LEL
CO
2
ppm
O
2
Vol%
CO
ppm
H
S
2
ppm
NH
3
ppm
CH
4
VOL%
CO
2
ppm
O
2
Vol%
CO
ppm
H
S
2
ppm
NH
3
ppm
O
2
VOL%
O
2
Vol%
CO
ppm
H
S
2
ppm
NH
3
ppm

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents