7 7 .2 C C leaning
CAUTION
Malfunctions from cleaning in an ultrasonic unit.
Defects in the product.
▶ Only clean manually or in a thermodisinfector.
7.2.1 C C leaning: Manual cleaning - external
Accessories required:
▪ Tap water 30 °C ± 5 °C (86 °F ± 10 °F)
▪ Brush, e.g. medium-hard toothbrush
Brush off under flowing tap water.
7.2.2 C C leaning: Automated external cleaning
KaVo recommends thermodisinfectors in accordance with EN ISO 15883
that are operated with alkaline cleaning agents at a pH of max. 10 (e.g.
Miele G 7781 / G 7881 – validation was performed with the ""VARIO-TD"
programme, "neodisher
mediclean" cleaning agent, "neodisher
®
tralisation agent, and "neodisher
to the compatibility of materials with respect to KaVo products).
For program settings as well as cleansers and disinfectants to be used,
please refer to the Instructions for Use of the thermodisinfector (com‐
plying with max. pH value of 10).
In order to prevent negative effects on the medical device, make sure
that the interior and the exterior of the medical device are dry, and then
lubricate immediately with care agents from the KaVo care system.
7 7 .2.3 C C leaning: Manual cleaning - internal
Can only be done with KaVo CLEANspray or KaVo DRYspray.
Cover the medical device with the KaVo CLEANpac bag, and place it
on the corresponding care adapter. Press the spray button three times
for 2 seconds each time. Remove the medical device from the spray
attachment and let the cleaner work for one minute.
Afterwards, rinse for 3-5 seconds with KaVo DRYspray.
S S ee also: KaVo CLEANspray / KaVo DRYspray Instructions for Use
Note
KaVo CLEANspray and KaVo DRYspray for manual interior cleaning are
only available in the following countries:
Germany, Austria, Switzerland, Italy, Spain, Portugal, France, Luxem‐
bourg, Belgium, Netherlands, United Kingdom, Denmark, Sweden, Fin‐
land and Norway.
In other countries, only automated interior cleaning with thermodisinfec‐
tors in accordance with ISO 15883-1.