6 Preparation methods according to ISO 17664 | 6.1 Preparations at the site of use

6 Preparation methods according to ISO 17664

Note
The following preparation procedure applies for the SONICflex instrument, torque
wrench and nozzle needle.

6.1 Preparations at the site of use

Hazard from nonsterile products.
There is a risk of infection from contaminated medical devices.
▶ Take suitable personal protective measures.
▶ Remove all residual cement, composite or blood without delay.
▶ Reprocess the medical device as soon as possible after treatment.
▶ The medical device must be dry when transported for reconditioning.
▶ Do not place it in a solution or similar.
▶ Remove the tip from the SONICflex using the torque wrench.

6.2 Cleaning

Malfunctions from cleaning in the ultrasonic unit.
Defects in the product.
▶ Only clean manually or in a thermodisinfector.

6.2.1 Manual cleaning - external

Accessories required:
▪ Tap water 30
C ± 5
o
▪ Brush, e.g. medium-hard toothbrush
▶ Brush it off under running tap water using for example a medium-hard toothbrush.
▶ Prepare the tip according to the manufacturer's instructions for use.

6.2.2 Automated external cleaning

KaVo recommends thermodisinfectors in accordance with EN ISO 15883-1, which are
operated with alkaline cleaning agents with a pH value of max. 10 (e.g. Miele G
7781 / G 7881 – Validation was carried out with Programme "VARIO-TD", cleaning
agent "neodisher® mediclean", neutralisation agent "neodisher® Z" and rinsing agent
"neodisher® mielclear" and only applies to the material compatibility with KaVo prod‐
ucts).
Instructions for use SONICflex 2000 N - 0.571.5020
WARNING
CAUTION
C (86
F ± 10
F)
o
o
o
15 / 26