Instructions for use KaVo polymerisation light Poly One
6 Reprocessing methods according to ISO 17664 | 6.1 Cleaning
6 Reprocessing methods according to ISO 17664
KaVo recommends to use original KaVo parts
since their safety, function, and specific suitability have been tested in extensive
tests.
CAUTION
Damage caused by improper cleaning/disinfection.
▶ Do not wash the curing light Poly One in the thermodisinfector.
▶ Do not sterilise the curing light Poly One in the hot-air steriliser.
▶ Do not place the curing light Poly One in a disinfection or ultrasound bath.
Note
For cleaning/reprocessing of the fibre-optic cable of the curing light Poly One, the
fibre-optic cable must be removed from the handpiece.
6.1 Cleaning
6.1.1 Manual cleaning
Note
Combine manual cleaning with disinfection.
Internal cleaning
Not applicable.
External cleaning
Accessories required:
▪ Tap water 30°C ± 5°C
▪ Brush, e.g. medium-hard toothbrush
for operation and repair only,
®
17/22