10.1.2 Test Devlces
Item
1001
1
Testflowmeter
, 6-50 Lj
min (Suction FlowTest)
NotShown
10-4
Low-pressure Leak Test Device
Low-pressure Leak Test Device
Low-pressure Leak Test Device
Flow test device capable of measuring 0-15 L/min
with an accuracy
of 士 2%
of reading
Vacuum test gauge capable of measuring 0 to 550 mm Hg
with an accuracy
of 士 1
% of reading
Test device capable of measuring 0-30 L/min (see Item 1 above)
Leakage current test device
Test device capable of measuring 689 kPa (100 psi)
Low-pressure test device
(dψtal
manometer ortest gauge)
with an accuracy
of 士 2%
of reading
Test Devlces
1
Stock Number
1006-8431-000
(negative pressure)
0309-1319-800
(positive pressure -ISO)
1001-8976-000
(positive pressure - BSI)
1001-8975-000
Referto section 6.5
Referto section 6.6
Referto section 6.6
Referto section 3.15
Referto section 5.1
Referto section 5.6.2