Aespire 7100 Planned Maintenance - GE Aespire 7100 Technical Reference Manual

Anestesia machine
Table of Contents

Advertisement

Aespire 7100
6.2 Aespire 7100 Planned Maintenance
Serial Number:
Hospital:
口 12
months
口 24
month
Date: 阳'/MM/DD)
performed by:
/
/
而 is
section covers the regular maintenance procedures (minimum
requirements) needed to make sure that the Aespire 7100 anesthesia
machine - including the ventilator - operates to specifications.
6.2.1
Every 阳'elve
(12)
Perform the following steps every 12 months.
months
For details , refer to the sections listed.
Parts Replacement
Replace
the vaporizer port o-rings (Section 9.8.1)
(Kit Stock Number 1102-3016-000)
Machlne Check5 and Te5ts
Refer to the Aesplre User's Reference Manual
,
Part 2.
Perform the following steps:
口1.
User maintenance listed below. Including disassembly , inspection ,
cleaning and parts replacement as required (Section 3 and Section 2).
• AGSS Maintenance:
Empty any condensate from the reservoir (disposable item).
Inspect air brake for occlusion.
Inspect , clean or replace filter on active AGSS.
• Breathing Circuit Maintenance
• Bellows Assembly Maintenance
• Bellows AssemblyTests
• O
2
Sensor Calibration
• Flow Sensor Calibration
Refer to IIsted sectlons
In 协Ismanual.
Perform the following steps:
口1.
MOPV
pre臼U 陀 relief
valve test (Section 6.3)
6-4
08/07 Mll10140