9.18 Replace task light and switch
9-48
9.18.1 To replace
the task- lI ght
swltch
9.18.2 To replace
the task- lI ght
clrcult board
Remove the four screws (A) that hold the task-light lens to the upper shel f.
D
B
c
1. Using a small needle-nose
plie邸,
disconnect the switch harness from the task-light
circuit board connector (8).
2. Remove the two screws (C) that hold the switch retainer plate to the upper shelf.
3. Transferthe switch retainer plate to the new switch , counter-sunkside to the outside.
4. Mount the switch to the upper shelf.
5. Remount the task-light assembly. Ensure that the switch harness and the task-light
harness wires are positioned in their respective recesses and are not pinched under
the task-light lens.
1. Using a small needle-nose
plie邸,
• disconnect the switch harness from the task light circuit board connector (8).
• disconnect the task-light harness from the task light circuit board connector (D).
2. Slide the task-light circuit board out of the lens.
3. Slide the new task light into the lens , ensuring that the connectors are aligned with
the lens cutouts.
4. Plug the task-light harness and the switch hamess into their respective connectors
on the task-light circuit board. Use a small screwdriver to push the connectors
securely into place.
5. Remount the task-light assembly. Ensure that the switch hamess and the task-light
hamess wires are positioned in their respective recesses and are not pinched under
the task-light lens.
08/07 M1110140