GE Aespire 7100 Technical Reference Manual page 204

Anestesia machine
Table of Contents

Advertisement

Aespire
7100
Message
A1
arm/Error
Cause
User Actlon/Concems
Servlce Repalr
PresjVol Mon
UserAlarm
ACGO is set to auxiliary gas
Connect the patient ci rcuit to the
Inactive
outlet
auxiliary gas outlet or set the
switch to the common gas
outlet for normal operation.
Replace O
2
Sensor
UserAlarm
O
2
%<5%
Make sure patient receives O
2 .
If persists , check:
Does the sensor see 21 % O
2
i n
• O
2
sensor cable
room ai r?
• VMB
Use different monitor.
• Cable to CBjJ12
Calibrate O
2
sensor.
• Control board
Replace O
2
sensor.
Service Calibration
UserAlarm
Internal calibrations are
刊 e
system is operationa l.
Service Mode:
A
necessary for maximum
• Calibrations
accuracy.
Select Gas Outlet
UserAlarm
Fresh gas may notflowtothe Selectthe common gas outlet or
patien
t.
connectthe patient circuitto
ACGO is On , but flow sensors
theACGO.
have seen 3 breaths in
Note: Th e bag arm will not
patient circuit during the
ventilate a patient at the
last 30 seconds.
auxiliary outle
t.
7-38
08/07 Mll10140