GE Aespire 7100 Technical Reference Manual page 228

Anestesia machine
Table of Contents

Advertisement

Aespire 7100
9-10
9.5.5 Replace
3rd-gas cyllnder
supply module
1. Remove the lower rear cover (Section 9.2.2).
2. Disconnect the high-pressure cylinder gauge fitting.
3. Disconnect the output tube
fi忧ing.
4. Remove the three mounting screws and lockwashers.
Outlet tube flttlng
5. To reassemble
,
perform the previous steps in reverse order.
• Tighten the high-pressure fitting as detailed in Section 9.5.1
• Pull on the cylinder output fitting to ensure it is locked in place.
6. Check the output ofthe regulator BEFORE you install the rear cover. Ad just if
nece臼ary
(Section 5.1).
7. Perform the checkout procedure (Section 3).
08/07 M1110140